Intsructions:
- Use the G-drill to remove the dentin until the tip of the G-drill is close to or touches the surface of the broken file.
- Select the appropriate size of guide cobalt (80,90,110 size) and continue to remove the dentin channel.
- After cutting the dentin, according to the size of the broken file, choose a suitable size of drill, grind the file channel at low speed in reverse rotation (recommended speed 150 rpm)
- Using the cricothyroid sleeve to remove the dentin around the broken file by counter-rotating at low speed. Until the broken file is exposed 2-3m, if the root canal is too curved for the cricothyroid sleeve to penetrate deeply, use the guide drill (60-70 size) to remove the dentin around the broken file at low speed until the broken file is exposed 2-3mm.
- Install a suotable size of clamping sleeve.
- Use the cannula mouth to cover the broken file.
- Press the handle downward to make the clamping sleeve clamp the broken file (pay attention to the pressure in place), after the clamping sleeve is tightened (pay attention to the pressure). Press the handle hard and pull out the broken file.
- Sterilize after use.
Operating Environment
- Operating temperature 5 degree celsius to 40 degree celsius
- Operating humidity 20% RH - 80%RH
- Atmospheric pressure 86kPa - 106kPa
Storage and transportation
- Storage temperature: 10 degree celsius to 55 degree celsius
- Storage humidity ≤ 93% RH
- Atmospheric pressure 50kPa - 106kPa